Lead
A centrifuge is an instrument often used in biological experiments. It uses a high-speed rotation of a rotating shaft to generate centrifugal force to separate substances of different densities. It can achieve the purpose of initial separation and purification and is an indispensable instrument in the laboratory. Laboratory accidents caused by improper use of the centrifuge in the laboratory also occur from time to time.
1
Cause of the accident
The main reasons for analyzing the accident are as follows:
1. When the centrifuge tube is balanced, the balance of the centrifuge tube is more than 0.1 to 1 g due to the long-term use of the balance. The weighing error of the sample, the overflow of the centrifuged sample liquid during centrifugation and the corrosion of the rotor are the main factors causing the rotor imbalance.
2. The high vacuum state in the centrifugal chamber during the process causes the centrifuge tube to rupture and the sample liquid to overflow, causing the rotor to be unbalanced and causing shaft bending or shaft breakage.
3. When using a horizontal rotor, the serial number of the gondola and the serial number of the horizontal rotor main body are not carefully read, which affects the dynamic balance of the rotor.
4. The rubber seal ring and the rotor cap rubber seal ring in the centrifuge cap are improperly used and improperly disinfected, such as long-term internal cracking and high-temperature disinfection (using dry box baking), etc., aging cracking, loss of sealing effect, etc. Overflow at high speeds, allowing the rotor to operate in an unbalanced state.
5. Scope and disinfection methods, using chemical solvents and improper disinfectant treatment, causing the centrifuge tube to swell and rupture during operation and cause an accident.
6. Due to carelessness in the work, the rotor cover is not tightened or the rotor cover and the rotor handle are interchanged, causing the screw to not match. When the machine is running, the rotor cover is thrown and a serious shaft breakage occurs.
7. The centrifuge tube is aged or the high-speed centrifuge centrifuge tube is used on the ultracentrifuge to break the centrifuge tube during operation, and the sample liquid overflows to cause an imbalance and bend the shaft.
2
Treatment
When a centrifuge tube breaks, it should:
1. Turn off the power and keep the centrifuge lid closed for 30 minutes. If the rupture occurs after the centrifuge has stopped running, immediately close the cover for 30 minutes.
2. When cleaning, put a pair of thick rubber gloves on the outside of the disposable gloves, and use the tweezers or the cotton tweezers to take the pieces. All the broken test tubes, crushing barrels and casings should be placed in the sharp box, according to the waste disposal regulations. Process it. Unbroken covered tubes disinfect the surface of contaminated containers with chlorine-containing disinfectant.
3. The inner surface of the centrifuge can be wiped twice with a chlorine-containing disinfectant or 75% alcohol with an effective chlorine concentration of 0.5%, then wiped with water and dried.
3
Precaution
Once the centrifuge fails, the aerosol may leak into the open environment, causing potential contamination. As can be seen from the figure below, the centrifuge centrifuge tube rupture is a severe aerosol leak, in order to prevent aerosols caused by the use of centrifuges. Leakage, in addition to following the correct operation of the centrifuge, you can use the centrifuge cover and use the centrifuge exhaust fan.
4
Products Recommended
Versati TM Centrifuge
The best choice for high performance and superior quality benchtop centrifuges
Esco is proud to introduce our Versati TM centrifuges with maintenance-free motors, powerful mechanical properties and an intelligent Ecourage TM microcomputer processor control system for maximum reliability and safety. Microcentrifuges and low-speed to high-speed multi-purpose centrifuges are available in a variety of models with rotors, adapters and other accessories. Versati TM has strong versatility, a variety of features to meet the needs of most applications and with a variety of test tubes, microplate header and the like to match the laboratory supplies.
Features:
Genuine-Protec TM Safety Cover Design
Motor drive locks ensure safety and reliability.
When the rotor stops at a speed of 0, the cover can be opened.
The electromagnetically controlled cover runs automatically and smoothly for one-handed operation.
V-balance TM rotor imbalance detection system
Equipped with automatic imbalance detection system. When the rotor detects a weight imbalance, it stops accelerating and the rotor decelerates until it stops.
Rotor Smartdrive TM Automatic Identification System
The replacement rotors are automatically identified and the speed of each rotor can only be set and operated at the maximum speed allowed by the rotor to ensure a longer motor system life.
Variety of rotors available
Powerful motors provide high centrifugal force, MCV / MCR and TCV up to 21380g and TCR up to 24381g.
Up to 7 rotors are available for the MCV/MCR; up to 11 rotors are available for the TCV/TCR to meet the user's experimental needs.
Inner cavity integrated design
The side walls and bottom plate of the centrifuge chamber are formed from a single piece of steel. The inner cavity has no dead angle design and is easy to clean, which can reduce the possibility of pollution.
Maintenance free, brushless motor
Made in Germany, no need to replace bearings, no need to change lubricants.
No carbon brush, low interference. No power sparks, no interference to radio equipment.
The superior mechanical design makes it run smoothly with low noise.
Extend life and save maintenance costs.
Emergency switch
In the event of a power outage or instrument failure, the centrifuge cover can be manually opened by an emergency switch to protect your sample.
Esco Versati TM centrifuges desktop centrifuge ultimate choice, excellent performance, stable quality, to provide you with the ultimate in reliability and security!
5
Conclusion
After reading this article, do you have any understanding of the precautions and accident handling measures used in the centrifuge? Experimental safety is a top priority, and Esco sincerely hopes that every user can safely and efficiently use the centrifuge!
If you would like to know more about centrifuges, you can visit our official website to find out more.